Concept URI | https://kodlistor.miljodatasamverkan.se/def/vocabulary/DATA_MATVARDE/ANALYS_MET/IVL_TENNORG |
---|---|
Preferred label | IVL_TENNORG |
Definition | IVL-Tennorganiska - A six point calibration curve was prepared by diluting Organotin mix 8 stock solution (LGC Promochem) to which fresh solutions of monophenyltin trichloride and diphenyltin dichloride in methanol had been added. After ethylation using sodium tetraethylborate water was added and the ethylates extracted with hexane. Instrumental analysis was carried out using a 7890A gas chromatograph connected to a 7010 triple quadrupole mass spectrometer (Agilent) used in electron ionization and multiple reaction monitoring (MRM) mode. For most compounds two MRMs were recorded, one as a quantifier and the other as a qualifier to increase specificity |