Concept URI | https://kodlistor.miljodatasamverkan.se/def/vocabulary/DATA_MATVARDE/ANALYS_MET/IVL_FLOT |
---|---|
Preferred label | IVL_FLOT |
Definition | IVL method IVL-FLOT - Determination of flotation chemicals used in the mining industry was carried out using a liquid chromatography-mass spectrometer (LC-MS/MS). Sample volume was extracted using solid-phase extraction (SPE) columns with a WAX phase (Cleanert PWAX, Phenomenex, WA1506), then eluted with 5/95 (v/v) ammonia (25%)/methanol, and analyzed using an Acquity UPLC BEH C18 column. Final analysis was performed on a Shimadzu LCMS-8050 system, including two gradient pumps (LC-40D x3), an autosampler (SIL-40C X3), and a column oven (CTO-40C). The flow rate was set to 0.3 mL/min, and the injection volume was 4 µL. The gradient program consisted of A: MQ-water with 0.12% ammonia (25%) and B: ACN |
Notation | |
Status | Valid |
Status Modified | 2025-08-25 |
Accepted Date | 2025-08-25 |
Not Accepted Date |